5 Development of Multi-Leaf Collimator-Based Isocenter Verification Tool Using Electrical Portal Imaging Device for Stereotactic Radiosurgery
Authors: Panatda Intanin, Sangutid Thongsawad, Chirapha Tannanonta, Todsaporn Fuangrod
Abstract:
Stereotactic radiosurgery (SRS) is a highly precision delivery technique that requires comprehensive quality assurance (QA) tests prior to treatment delivery. An isocenter of delivery beam plays a critical role that affect the treatment accuracy. The uncertainty of isocenter is traditionally accessed using circular cone equipment, Winston-Lutz (WL) phantom and film. This technique is considered time consuming and highly dependent on the observer. In this work, the development of multileaf collimator (MLC)-based isocenter verification tool using electronic portal imaging device (EPID) was proposed and evaluated. A mechanical isocenter alignment with ball bearing diameter 5 mm and circular cone diameter 10 mm fixed to gantry head defines the radiation field was set as the conventional WL test method. The conventional setup was to compare to the proposed setup; using MLC (10 x 10 mm) to define the radiation filed instead of cone. This represents more realistic delivery field than using circular cone equipment. The acquisition from electronic portal imaging device (EPID) and radiographic film were performed in both experiments. The gantry angles were set as following: 0°, 90°, 180° and 270°. A software tool was in-house developed using MATLAB/SIMULINK programming to determine the centroid of radiation field and shadow of WL phantom automatically. This presents higher accuracy than manual measurement. The deviation between centroid of both cone-based and MLC-based WL tests were quantified. To compare between film and EPID image, the deviation for all gantry angle was 0.26±0.19mm and 0.43±0.30 for cone-based and MLC-based WL tests. For the absolute deviation calculation on EPID images between cone and MLC-based WL test was 0.59±0.28 mm and the absolute deviation on film images was 0.14±0.13 mm. Therefore, the MLC-based isocenter verification using EPID present high sensitivity tool for SRS QA.Keywords: isocenter verification, quality assurance, EPID, SRS
Procedia PDF Downloads 1524 Quality Assurance Comparison of Map Check 2, Epid, and Gafchromic® EBT3 Film for IMRT Treatment Planning
Authors: Khalid Iqbal, Saima Altaf, M. Akram, Muhammad Abdur Rafaye, Saeed Ahmad Buzdar
Abstract:
Objective: Verification of patient-specific intensity modulated radiation therapy (IMRT) plans using different 2-D detectors has become increasingly popular due to their ease of use and immediate readout of the results. The purpose of this study was to test and compare various 2-D detectors for dosimetric quality assurance (QA) of intensity-modulated radiotherapy (IMRT) with the vision to find alternative QA methods. Material and Methods: Twenty IMRT patients (12 of brain and 8 of the prostate) were planned on Eclipse treatment planning system using Varian Clinac DHX on both energies 6MV and 15MV. Verification plans of all such patients were also made and delivered to Map check2, EPID (Electronic portal imaging device) and Gafchromic EBT3. Gamma index analyses were performed using different criteria to evaluate and compare the dosimetric results. Results: Statistical analysis shows the passing rate of 99.55%, 97.23% and 92.9% for 6MV and 99.53%, 98.3% and 94.85% for 15 MV energy using a criteria of ±5% of 3mm, ±3% of 3mm and ±3% of 2mm respectively for brain, whereas using ±5% of 3mm and ±3% of 3mm gamma evaluation criteria, the passing rate is 94.55% and 90.45% for 6MV and 95.25%9 and 95% for 15 MV energy for the case of prostate using EBT3 film. Map check 2 results shows the passing rates of 98.17%, 97.68% and 86.78% for 6MV energy and 94.87%,97.46% and 88.31% for 15 MV energy respectively for brain using a criteria of ±5% of 3mm, ±3% of 3mm and ±3% of 2mm, whereas using ±5% of 3mm and ±3% of 3mm gamma evaluation criteria gives the passing rate of 97.7% and 96.4% for 6MV and 98.75%9 and 98.05% for 15 MV energy for the case of prostate. EPID 6 MV and gamma analysis shows the passing rate of 99.56%, 98.63% and 98.4% for the brain, 100% and 99.9% for prostate using the same criteria as for map check 2 and EBT 3 film. Conclusion: The results demonstrate excellent passing rates were obtained for all dosimeter when compared with the planar dose distributions for 6 MV IMRT fields as well as for 15 MV. EPID results are better than EBT3 films and map check 2 because it is likely that part of this difference is real, and part is due to manhandling and different treatment set up verification which contributes dose distribution difference. Overall all three dosimeter exhibits results within limits according to AAPM report.120.Keywords: gafchromic EBT3, radiochromic film dosimetry, IMRT verification, EPID
Procedia PDF Downloads 4213 A Prospective Study of a Clinically Significant Anatomical Change in Head and Neck Intensity-Modulated Radiation Therapy Using Transit Electronic Portal Imaging Device Images
Authors: Wilai Masanga, Chirapha Tannanonta, Sangutid Thongsawad, Sasikarn Chamchod, Todsaporn Fuangrod
Abstract:
The major factors of radiotherapy for head and neck (HN) cancers include patient’s anatomical changes and tumour shrinkage. These changes can significantly affect the planned dose distribution that causes the treatment plan deterioration. A measured transit EPID images compared to a predicted EPID images using gamma analysis has been clinically implemented to verify the dose accuracy as part of adaptive radiotherapy protocol. However, a global gamma analysis dose not sensitive to some critical organ changes as the entire treatment field is compared. The objective of this feasibility study is to evaluate the dosimetric response to patient anatomical changes during the treatment course in HN IMRT (Head and Neck Intensity-Modulated Radiation Therapy) using a novel comparison method; organ-of-interest gamma analysis. This method provides more sensitive to specific organ change detection. Random replanned 5 HN IMRT patients with causes of tumour shrinkage and patient weight loss that critically affect to the parotid size changes were selected and evaluated its transit dosimetry. A comprehensive physics-based model was used to generate a series of predicted transit EPID images for each gantry angle from original computed tomography (CT) and replan CT datasets. The patient structures; including left and right parotid, spinal cord, and planning target volume (PTV56) were projected to EPID level. The agreement between the transit images generated from original CT and replanned CT was quantified using gamma analysis with 3%, 3mm criteria. Moreover, only gamma pass-rate is calculated within each projected structure. The gamma pass-rate in right parotid and PTV56 between predicted transit of original CT and replan CT were 42.8%( ± 17.2%) and 54.7%( ± 21.5%). The gamma pass-rate for other projected organs were greater than 80%. Additionally, the results of organ-of-interest gamma analysis were compared with 3-dimensional cone-beam computed tomography (3D-CBCT) and the rational of replan by radiation oncologists. It showed that using only registration of 3D-CBCT to original CT does not provide the dosimetric impact of anatomical changes. Using transit EPID images with organ-of-interest gamma analysis can provide additional information for treatment plan suitability assessment.Keywords: re-plan, anatomical change, transit electronic portal imaging device, EPID, head, and neck
Procedia PDF Downloads 2162 The Use of the Matlab Software as the Best Way to Recognize Penumbra Region in Radiotherapy
Authors: Alireza Shayegan, Morteza Amirabadi
Abstract:
The y tool was developed to quantitatively compare dose distributions, either measured or calculated. Before computing ɣ, the dose and distance scales of the two distributions, referred to as evaluated and reference, are re-normalized by dose and distance criteria, respectively. The re-normalization allows the dose distribution comparison to be conducted simultaneously along dose and distance axes. Several two-dimensional images were acquired using a Scanning Liquid Ionization Chamber EPID and Extended Dose Range (EDR2) films for regular and irregular radiation fields. The raw images were then converted into two-dimensional dose maps. Transitional and rotational manipulations were performed for images using Matlab software. As evaluated dose distribution maps, they were then compared with the corresponding original dose maps as the reference dose maps.Keywords: energetic electron, gamma function, penumbra, Matlab software
Procedia PDF Downloads 3011 Linac Quality Controls Using An Electronic Portal Imaging Device
Authors: Domingo Planes Meseguer, Raffaele Danilo Esposito, Maria Del Pilar Dorado Rodriguez
Abstract:
Monthly quality control checks for a Radiation Therapy Linac may be performed is a simple and efficient way once they have been standardized and protocolized. On the other hand this checks, in spite of being imperatives, require a not negligible execution times in terms of machine time and operators time. Besides it must be taken into account the amount of disposable material which may be needed together with the use of commercial software for their performing. With the aim of optimizing and standardizing mechanical-geometric checks and multi leaves collimator checks, we decided to implement a protocol which makes use of the Electronic Portal Imaging Device (EPID) available on our Linacs. The user is step by step guided by the software during the whole procedure. Acquired images are automatically analyzed by our programs all of them written using only free software.Keywords: quality control checks, linac, radiation oncology, medical physics, free software
Procedia PDF Downloads 199