Search results for: gafchromic EBT3
Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 13

Search results for: gafchromic EBT3

13 Quality Assurance Comparison of Map Check 2, Epid, and Gafchromic® EBT3 Film for IMRT Treatment Planning

Authors: Khalid Iqbal, Saima Altaf, M. Akram, Muhammad Abdur Rafaye, Saeed Ahmad Buzdar

Abstract:

Objective: Verification of patient-specific intensity modulated radiation therapy (IMRT) plans using different 2-D detectors has become increasingly popular due to their ease of use and immediate readout of the results. The purpose of this study was to test and compare various 2-D detectors for dosimetric quality assurance (QA) of intensity-modulated radiotherapy (IMRT) with the vision to find alternative QA methods. Material and Methods: Twenty IMRT patients (12 of brain and 8 of the prostate) were planned on Eclipse treatment planning system using Varian Clinac DHX on both energies 6MV and 15MV. Verification plans of all such patients were also made and delivered to Map check2, EPID (Electronic portal imaging device) and Gafchromic EBT3. Gamma index analyses were performed using different criteria to evaluate and compare the dosimetric results. Results: Statistical analysis shows the passing rate of 99.55%, 97.23% and 92.9% for 6MV and 99.53%, 98.3% and 94.85% for 15 MV energy using a criteria of ±5% of 3mm, ±3% of 3mm and ±3% of 2mm respectively for brain, whereas using ±5% of 3mm and ±3% of 3mm gamma evaluation criteria, the passing rate is 94.55% and 90.45% for 6MV and 95.25%9 and 95% for 15 MV energy for the case of prostate using EBT3 film. Map check 2 results shows the passing rates of 98.17%, 97.68% and 86.78% for 6MV energy and 94.87%,97.46% and 88.31% for 15 MV energy respectively for brain using a criteria of ±5% of 3mm, ±3% of 3mm and ±3% of 2mm, whereas using ±5% of 3mm and ±3% of 3mm gamma evaluation criteria gives the passing rate of 97.7% and 96.4% for 6MV and 98.75%9 and 98.05% for 15 MV energy for the case of prostate. EPID 6 MV and gamma analysis shows the passing rate of 99.56%, 98.63% and 98.4% for the brain, 100% and 99.9% for prostate using the same criteria as for map check 2 and EBT 3 film. Conclusion: The results demonstrate excellent passing rates were obtained for all dosimeter when compared with the planar dose distributions for 6 MV IMRT fields as well as for 15 MV. EPID results are better than EBT3 films and map check 2 because it is likely that part of this difference is real, and part is due to manhandling and different treatment set up verification which contributes dose distribution difference. Overall all three dosimeter exhibits results within limits according to AAPM report.120.

Keywords: gafchromic EBT3, radiochromic film dosimetry, IMRT verification, EPID

Procedia PDF Downloads 397
12 Measurement of IMRT Dose Distribution in Rando Head and Neck Phantom using EBT3 Film

Authors: Pegah Safavi, Mehdi Zehtabian, Mohammad Amin Mosleh-Shirazi

Abstract:

Cancer is one of the leading causes of death in the world. Radiation therapy is one of the main choices for cancer treatment. Intensity-modulated radiation therapy is a new type of radiation therapy technique available for vital structures such as the parathyroid glands. It is very important to check the accuracy of the delivered IMRT treatment because any mistake may lead to more complications for the patient. This paper describes an experiment to determine the accuracy of a dose measured by EBT3 film. To test this method, the EBT3 film on the head and neck of the Rando phantom was irradiated by an IMRT device and the irradiation was repeated twice. Finally, the dose designed by the irradiation system was compared with the dose measured by the EBT3 film. Using this criterion, the accuracy of the EBT3 film was evaluated. When using this criterion, a 95% agreement was reached between the planned treatment and the measured values.

Keywords: EBT3, phantom, accuracy, cancer, IMRT

Procedia PDF Downloads 117
11 Viability of EBT3 Film in Small Dimensions to Be Use for in-Vivo Dosimetry in Radiation Therapy

Authors: Abdul Qadir Jangda, Khadija Mariam, Usman Ahmed, Sharib Ahmed

Abstract:

The Gafchromic EBT3 film has the characteristic of high spatial resolution, weak energy dependence and near tissue equivalence which makes them viable to be used for in-vivo dosimetry in External Beam and Brachytherapy applications. The aim of this study is to assess the smallest film dimension that may be feasible for the use in in-vivo dosimetry. To evaluate the viability, the film sizes from 3 x 3 mm to 20 x 20 mm were calibrated with 6 MV Photon and 6 MeV electron beams. The Gafchromic EBT3 (Lot no. A05151201, Make: ISP) film was cut into five different sizes in order to establish the relationship between absorbed dose vs. film dimensions. The film dimension were 3 x 3, 5 x 5, 10 x 10, 15 x 15, and 20 x 20 mm. The films were irradiated on Varian Clinac® 2100C linear accelerator for dose range from 0 to 1000 cGy using PTW solid water phantom. The irradiation was performed as per clinical absolute dose rate calibratin setup, i.e. 100 cm SAD, 5.0 cm depth and field size of 10x10 cm2 and 100 cm SSD, 1.4 cm depth and 15x15 cm2 applicator for photon and electron respectively. The irradiated films were scanned with the landscape orientation and a post development time of 48 hours (minimum). Film scanning accomplished using Epson Expression 10000 XL Flatbed Scanner and quantitative analysis carried out with ImageJ freeware software. Results show that the dose variation with different film dimension ranging from 3 x 3 mm to 20 x 20 mm is very minimal with a maximum standard deviation of 0.0058 in Optical Density for a dose level of 3000 cGy and the the standard deviation increases with the increase in dose level. So the precaution must be taken while using the small dimension films for higher doses. Analysis shows that there is insignificant variation in the absorbed dose with a change in film dimension of EBT3 film. Study concludes that the film dimension upto 3 x 3 mm can safely be used up to a dose level of 3000 cGy without the need of recalibration for particular dimension in use for dosimetric application. However, for higher dose levels, one may need to calibrate the films for a particular dimension in use for higher accuracy. It was also noticed that the crystalline structure of the film got damage at the edges while cutting the film, which can contribute to the wrong dose if the region of interest includes the damage area of the film

Keywords: external beam radiotherapy, film calibration, film dosimetery, in-vivo dosimetery

Procedia PDF Downloads 462
10 Evaluation of Uniformity for Gafchromic Sheets for Film Dosimetry

Authors: Fayzan Ahmed, Saad Bin Saeed, Abdul Qadir Jangda

Abstract:

Gafchromic™ sheet are extensively used for the QA of intensity modulated radiation therapy and other in-vivo dosimetry. Intra-sheet Non-uniformity of scanner as well as film causes undesirable fluctuations which are reflected in dosimetry The aim of this study is to define a systematic and robust method to investigate the intra-sheet uniformity of the unexposed Gafchromic Sheets and the region of interest (ROI) of the scanner. Sheets of lot No#: A05151201 were scanned before and after the expiry period with the EPSON™ XL10000 scanner in the transmission mode, landscape orientation and 72 dpi resolution. ROI of (8’x 10’ inches) equal to the sheet dimension in the center of the scanner is used to acquire images with full transmission, block transmission and with sheets in place. 500 virtual grids, created in MATALB® are imported as a macros in ImageJ (1.49m Wayne Rasband) to analyze the images. In order to remove the edge effects, the outer 86 grids are excluded from the analysis. The standard deviation of the block transmission and full transmission are 0.38% and 0.66% confirming a higher uniformity of the scanner. Expired and non-expired sheets have standard deviations of 2.18% and 1.29%, show that uniformity decreases after expiry. The results are promising and indicates a good potential of this method to be used as a uniformity check for scanner and unexposed Gafchromic sheets.

Keywords: IMRT, film dosimetry, virtual grids, uniformity

Procedia PDF Downloads 453
9 A Simple Method for Evaluation of Uniformity for Gafchromic Sheets for Film Dosimetry

Authors: Fayzan Ahmed, Saad Bin Saeed, Abdul Qadir Jangda

Abstract:

Gafchromic™ sheet are extensively used for the QA of intensity modulated radiation therapy and other in-vivo dosimetry. Intra-sheet Non-uniformity of scanner as well as film causes undesirable fluctuations which are reflected in dosimetry The aim of this study is to define a systematic and robust method to investigate the intra-sheet uniformity of the unexposed Gafchromic Sheets and the region of interest (ROI) of the scanner. Sheets of lot No#: A05151201 were scanned before and after the expiry period with the EPSON™ XL10000 scanner in the transmission mode, landscape orientation, and 72 dpi resolution. ROI of (8’x 10’ inches) equal to the sheet dimension in the center of the scanner is used to acquire images with full transmission, block transmission and with sheets in place. 500 virtual grids, created in MATALB® are imported as a macros in ImageJ (1.49m Wayne Rasband) to analyze the images. In order to remove the edge effects, the outer 86 grids are excluded from the analysis. The standard deviation of the block transmission and full transmission are 0.38% and 0.66% confirming a higher uniformity of the scanner. Expired and non-expired sheets have standard deviations of 2.18% and 1.29%, show that uniformity decreases after expiry. The results are promising and indicate a good potential of this method to be used as a uniformity check for scanner and unexposed Gafchromic sheets.

Keywords: IMRT, film dosimetry, virtual grids, uniformity

Procedia PDF Downloads 398
8 Measurement and Analysis of Radiation Doses to Radiosensitive Organs from CT Examination of the Cervical Spine Using Radiochromic Films and Monte Carlo Simulation Based Software

Authors: Khaled Soliman, Abdullah Alrushoud, Abdulrahman Alkhalifah, Raed Albathi, Salman Altymiat

Abstract:

Radiation dose received by patients undergoing Computed Tomography (CT) examination of the cervical spine was evaluated using Gafchromic XR-QA2 films and CT-EXPO software (ver. 2.3), in order to document our clinical dose values and to compare our results with other benchmarks reported in the current literature. Radiochromic films were recently used as practical dosimetry tool that provides dose profile information not available using the standard ionisation chamber routinely used in CT dosimetry. We have developed an in-house program to use the films in order to calculate the Entrance Dose Length Product (EDLP) in (mGy.cm) and to relate the EDLP to various organ doses calculated using the CT-EXPO software. We also calculated conversion factor in (mSv/mGy.cm) relating the EDLP to the effective dose (ED) from the examination using CT-EXPO software. Variability among different types of CT scanners and dose modulation methods are reported from at least three major CT brands available at our medical institution. Our work describes the dosimetry method and results are reported. The method can be used as in-vivo dosimetry method. But this work only reports results obtained from adult female anthropomorphic Phantom studies.

Keywords: CT dosimetry, gafchromic films, XR-QA2, CT-Expo software

Procedia PDF Downloads 440
7 Approaching In vivo Dosimetry for Kilovoltage X-Ray Radiotherapy

Authors: Rodolfo Alfonso, David Alonso, Albin Garcia, Jose Luis Alonso

Abstract:

Recently a new kilovoltage radiotherapy unit model Xstrahl 200 - donated to the INOR´s Department of Radiotherapy (DR-INOR) in the framework of a IAEA's technical cooperation project- has been commissioned. This unit is able to treat shallow and low deep laying lesions, as it provides 8 discrete beam qualities, from 40 to 200 kV. As part of the patient-specific quality assurance program established at DR-INOR for external beam radiotherapy, it has been recommended to implement in vivo dose measurements (IVD), as they allow effectively discovering eventual errors or failures in the radiotherapy process. For that purpose a radio-photoluminescence (RPL) dosimetry system, model XXX, -also donated to DR-INOR by the same IAEA project- has been studied and commissioned. Main dosimetric parameters of the RPL system, such as reproducibility, linearity, and filed size influence were assessed. In a similar way, the response of radiochromic EBT3 type film was investigated for purposes of IVD. Both systems were calibrated in terms of entrance surface dose. Results of the dosimetric commissioning of RPL and EBT3 for IVD, and their pre-clinical implementation through end-to-end test cases are presented. The RPL dosimetry seems more recommendable for hyper-fractionated schemes with larger fields and curved patient contours, as those in chest wall irradiations, where the use of more than one dosimeter could be required. The radiochromic system involves smaller corrections with field size, but it sensibility is lower; hence it is more adequate for hypo-fractionated treatments with smaller fields.

Keywords: glass dosimetry, in vivo dosimetry, kilovotage radiotherapy, radiochromic dosimetry

Procedia PDF Downloads 366
6 Comparative Evaluation of EBT3 Film Dosimetry Using Flat Bad Scanner, Densitometer and Spectrophotometer Methods and Its Applications in Radiotherapy

Authors: K. Khaerunnisa, D. Ryangga, S. A. Pawiro

Abstract:

Over the past few decades, film dosimetry has become a tool which is used in various radiotherapy modalities, either for clinical quality assurance (QA) or dose verification. The response of the film to irradiation is usually expressed in optical density (OD) or net optical density (netOD). While the film's response to radiation is not linear, then the use of film as a dosimeter must go through a calibration process. This study aimed to compare the function of the calibration curve of various measurement methods with various densitometer, using a flat bad scanner, point densitometer and spectrophotometer. For every response function, a radichromic film calibration curve is generated from each method by performing accuracy, precision and sensitivity analysis. netOD is obtained by measuring changes in the optical density (OD) of the film before irradiation and after irradiation when using a film scanner if it uses ImageJ to extract the pixel value of the film on the red channel of three channels (RGB), calculate the change in OD before and after irradiation when using a point densitometer, and calculate changes in absorbance before and after irradiation when using a spectrophotometer. the results showed that the three calibration methods gave readings with a netOD precision of doses below 3% for the uncertainty value of 1σ (one sigma). while the sensitivity of all three methods has the same trend in responding to film readings against radiation, it has a different magnitude of sensitivity. while the accuracy of the three methods provides readings below 3% for doses above 100 cGy and 200 cGy, but for doses below 100 cGy found above 3% when using point densitometers and spectrophotometers. when all three methods are used for clinical implementation, the results of the study show accuracy and precision below 2% for the use of scanners and spectrophotometers and above 3% for precision and accuracy when using point densitometers.

Keywords: Callibration Methods, Film Dosimetry EBT3, Flat Bad Scanner, Densitomete, Spectrophotometer

Procedia PDF Downloads 100
5 Skin-Dose Mapping for Patients Undergoing Interventional Radiology Procedures: Clinical Experimentations versus a Mathematical Model

Authors: Aya Al Masri, Stefaan Carpentier, Fabrice Leroy, Thibault Julien, Safoin Aktaou, Malorie Martin, Fouad Maaloul

Abstract:

Introduction: During an 'Interventional Radiology (IR)' procedure, the patient's skin-dose may become very high for a burn, necrosis and ulceration to appear. In order to prevent these deterministic effects, an accurate calculation of the patient skin-dose mapping is essential. For most machines, the 'Dose Area Product (DAP)' and fluoroscopy time are the only information available for the operator. These two parameters are a very poor indicator of the peak skin dose. We developed a mathematical model that reconstructs the magnitude (delivered dose), shape, and localization of each irradiation field on the patient skin. In case of critical dose exceeding, the system generates warning alerts. We present the results of its comparison with clinical studies. Materials and methods: Two series of comparison of the skin-dose mapping of our mathematical model with clinical studies were performed: 1. At a first time, clinical tests were performed on patient phantoms. Gafchromic films were placed on the table of the IR machine under of PMMA plates (thickness = 20 cm) that simulate the patient. After irradiation, the film darkening is proportional to the radiation dose received by the patient's back and reflects the shape of the X-ray field. After film scanning and analysis, the exact dose value can be obtained at each point of the mapping. Four experimentation were performed, constituting a total of 34 acquisition incidences including all possible exposure configurations. 2. At a second time, clinical trials were launched on real patients during real 'Chronic Total Occlusion (CTO)' procedures for a total of 80 cases. Gafchromic films were placed at the back of patients. We performed comparisons on the dose values, as well as the distribution, and the shape of irradiation fields between the skin dose mapping of our mathematical model and Gafchromic films. Results: The comparison between the dose values shows a difference less than 15%. Moreover, our model shows a very good geometric accuracy: all fields have the same shape, size and location (uncertainty < 5%). Conclusion: This study shows that our model is a reliable tool to warn physicians when a high radiation dose is reached. Thus, deterministic effects can be avoided.

Keywords: clinical experimentation, interventional radiology, mathematical model, patient's skin-dose mapping.

Procedia PDF Downloads 113
4 Comparison of an Anthropomorphic PRESAGE® Dosimeter and Radiochromic Film with a Commercial Radiation Treatment Planning System for Breast IMRT: A Feasibility Study

Authors: Khalid Iqbal

Abstract:

This work presents a comparison of an anthropomorphic PRESAGE® dosimeter and radiochromic film measurements with a commercial treatment planning system to determine the feasibility of PRESAGE® for 3D dosimetry in breast IMRT. An anthropomorphic PRESAGE® phantom was created in the shape of a breast phantom. A five-field IMRT plan was generated with a commercially available treatment planning system and delivered to the PRESAGE® phantom. The anthropomorphic PRESAGE® was scanned with the Duke midsized optical CT scanner (DMOS-RPC) and the OD distribution was converted to dose. Comparisons were performed between the dose distribution calculated with the Pinnacle3 treatment planning system, PRESAGE®, and EBT2 film measurements. DVHs, gamma maps, and line profiles were used to evaluate the agreement. Gamma map comparisons showed that Pinnacle3 agreed with PRESAGE® as greater than 95% of comparison points for the PTV passed a ± 3%/± 3 mm criterion when the outer 8 mm of phantom data were discluded. Edge artifacts were observed in the optical CT reconstruction, from the surface to approximately 8 mm depth. These artifacts resulted in dose differences between Pinnacle3 and PRESAGE® of up to 5% between the surface and a depth of 8 mm and decreased with increasing depth in the phantom. Line profile comparisons between all three independent measurements yielded a maximum difference of 2% within the central 80% of the field width. For the breast IMRT plan studied, the Pinnacle3 calculations agreed with PRESAGE® measurements to within the ±3%/± 3 mm gamma criterion. This work demonstrates the feasibility of the PRESAGE® to be fashioned into anthropomorphic shape, and establishes the accuracy of Pinnacle3 for breast IMRT. Furthermore, these data have established the groundwork for future investigations into 3D dosimetry with more complex anthropomorphic phantoms.

Keywords: 3D dosimetry, PRESAGE®, IMRT, QA, EBT2 GAFCHROMIC film

Procedia PDF Downloads 375
3 Reduction of the Risk of Secondary Cancer Induction Using VMAT for Head and Neck Cancer

Authors: Jalil ur Rehman, Ramesh C, Tailor, Isa Khan, Jahanzeeb Ashraf, Muhammad Afzal, Geofferry S. Ibbott

Abstract:

The purpose of this analysis is to estimate secondary cancer risks after VMAT compared to other modalities of head and neck radiotherapy (IMRT, 3DCRT). Computer tomography (CT) scans of Radiological Physics Center (RPC) head and neck phantom were acquired with CT scanner and exported via DICOM to the treatment planning system (TPS). Treatment planning was done using four arc (182-178 and 180-184, clockwise and anticlockwise) for volumetric modulated arc therapy (VMAT) , Nine fields (200, 240, 280, 320,0,40,80,120 and 160), which has been commonly used at MD Anderson Cancer Center Houston for intensity modulated radiation therapy (IMRT) and four fields for three dimensional radiation therapy (3DCRT) were used. True beam linear accelerator of 6MV photon energy was used for dose delivery, and dose calculation was done with CC convolution algorithm with prescription dose of 6.6 Gy. Primary Target Volume (PTV) coverage, mean and maximal doses, DVHs and volumes receiving more than 2 Gy and 3.8 Gy of OARs were calculated and compared. Absolute point dose and planar dose were measured with thermoluminescent dosimeters (TLDs) and GafChromic EBT2 film, respectively. Quality Assurance of VMAT and IMRT were performed by using ArcCHECK method with gamma index criteria of 3%/3mm dose difference to distance to agreement (DD/DTA). PTV coverage was found 90.80 %, 95.80 % and 95.82 % for 3DCRT, IMRT and VMAT respectively. VMAT delivered the lowest maximal doses to esophagus (2.3 Gy), brain (4.0 Gy) and thyroid (2.3 Gy) compared to all other studied techniques. In comparison, maximal doses for 3DCRT were found higher than VMAT for all studied OARs. Whereas, IMRT delivered maximal higher doses 26%, 5% and 26% for esophagus, normal brain and thyroid, respectively, compared to VMAT. It was noted that esophagus volume receiving more than 2 Gy was 3.6 % for VMAT, 23.6 % for IMRT and up to 100 % for 3DCRT. Good agreement was observed between measured doses and those calculated with TPS. The averages relative standard errors (RSE) of three deliveries within eight TLD capsule locations were, 0.9%, 0.8% and 0.6% for 3DCRT, IMRT and VMAT, respectively. The gamma analysis for all plans met the ±5%/3 mm criteria (over 90% passed) and results of QA were greater than 98%. The calculations for maximal doses and volumes of OARs suggest that the estimated risk of secondary cancer induction after VMAT is considerably lower than IMRT and 3DCRT.

Keywords: RPC, 3DCRT, IMRT, VMAT, EBT2 film, TLD

Procedia PDF Downloads 480
2 Dose Saving and Image Quality Evaluation for Computed Tomography Head Scanning with Eye Protection

Authors: Yuan-Hao Lee, Chia-Wei Lee, Ming-Fang Lin, Tzu-Huei Wu, Chih-Hsiang Ko, Wing P. Chan

Abstract:

Computed tomography (CT) scan of the head is a good method for investigating cranial lesions. However, radiation-induced oxidative stress can be accumulated in the eyes and promote carcinogenesis and cataract. In this regard, we aimed to protect the eyes with barium sulfate shield(s) during CT scans and investigate the resultant image quality and radiation dose to the eye. Patients who underwent health examinations were selectively enrolled in this study in compliance with the protocol approved by the Ethics Committee of the Joint Institutional Review Board at Taipei Medical University. Participants’ brains were scanned with a water-based marker simultaneously by a multislice CT scanner (SOMATON Definition Flash) under a fixed tube current-time setting or automatic tube current modulation (TCM). The lens dose was measured by Gafchromic films, whose dose response curve was previously fitted using thermoluminescent dosimeters, with or without barium sulfate or bismuth-antimony shield laid above. For the assessment of image quality CT images at slice planes that exhibit the interested regions on the zygomatic, orbital and nasal bones of the head phantom as well as the water-based marker were used for calculating the signal-to-noise and contrast-to-noise ratios. The application of barium sulfate and bismuth-antimony shields decreased 24% and 47% of the lens dose on average, respectively. Under topogram-based TCM, the dose saving power of bismuth-antimony shield was mitigated whereas that of barium sulfate shield was enhanced. On the other hand, the signal-to-noise and contrast-to-noise ratios of DSCT images were decreased separately by barium sulfate and bismuth-antimony shield, resulting in an overall reduction of the CNR. In contrast, the integration of topogram-based TCM elevated signal difference between the ROIs on the zygomatic bones and eyeballs while preferentially decreasing the signal-to-noise ratios upon the use of barium sulfate shield. The results of this study indicate that the balance between eye exposure and image quality can be optimized by combining eye shields with topogram-based TCM on the multislice scanner. Eye shielding could change the photon attenuation characteristics of tissues that are close to the shield. The application of both shields on eye protection hence is not recommended for seeking intraorbital lesions.

Keywords: computed tomography, barium sulfate shield, dose saving, image quality

Procedia PDF Downloads 241
1 Dosimetric Comparison among Different Head and Neck Radiotherapy Techniques Using PRESAGE™ Dosimeter

Authors: Jalil ur Rehman, Ramesh C. Tailor, Muhammad Isa Khan, Jahnzeeb Ashraf, Muhammad Afzal, Geofferry S. Ibbott

Abstract:

Purpose: The purpose of this analysis was to investigate dose distribution of different techniques (3D-CRT, IMRT and VMAT) of head and neck cancer using 3-dimensional dosimeter called PRESAGETM Dosimeter. Materials and Methods: Computer tomography (CT) scans of radiological physics center (RPC) head and neck anthropomorphic phantom with both RPC standard insert and PRESAGETM insert were acquired separated with Philipp’s CT scanner and both CT scans were exported via DICOM to the Pinnacle version 9.4 treatment planning system (TPS). Each plan was delivered twice to the RPC phantom first containing the RPC standard insert having TLD and film dosimeters and then again containing the Presage insert having 3-D dosimeter (PRESAGETM) by using a Varian True Beam linear accelerator. After irradiation, the standard insert including point dose measurements (TLD) and planar Gafchromic® EBT film measurement were read using RPC standard procedure. The 3D dose distribution from PRESAGETM was read out with the Duke Midsized optical scanner dedicated to RPC (DMOS-RPC). Dose volume histogram (DVH), mean and maximal doses for organs at risk were calculated and compared among each head and neck technique. The prescription dose was same for all head and neck radiotherapy techniques which was 6.60 Gy/friction. Beam profile comparison and gamma analysis were used to quantify agreements among film measurement, PRESAGETM measurement and calculated dose distribution. Quality assurances of all plans were performed by using ArcCHECK method. Results: VMAT delivered the lowest mean and maximum doses to organ at risk (spinal cord, parotid) than IMRT and 3DCRT. Such dose distribution was verified by absolute dose distribution using thermoluminescent dosimeter (TLD) system. The central axial, sagittal and coronal planes were evaluated using 2D gamma map criteria(± 5%/3 mm) and results were 99.82% (axial), 99.78% (sagital), 98.38% (coronal) for VMAT plan and found the agreement between PRESAGE and pinnacle was better than IMRT and 3D-CRT plan excludes a 7 mm rim at the edge of the dosimeter. Profile showed good agreement for all plans between film, PRESAGE and pinnacle and 3D gamma was performed for PTV and OARs, VMAT and 3DCRT endow with better agreement than IMRT. Conclusion: VMAT delivered lowered mean and maximal doses to organs at risk and better PTV coverage during head and neck radiotherapy. TLD, EBT film and PRESAGETM dosimeters suggest that VMAT was better for the treatment of head and neck cancer than IMRT and 3D-CRT.

Keywords: RPC, 3DCRT, IMRT, VMAT, EBT2 film, TLD, PRESAGETM

Procedia PDF Downloads 360