Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 33122
Design and Microfabrication of a High Throughput Thermal Cycling Platform with Various Annealing Temperatures
Authors: Sin J. Chen, Jyh J. Chen
Abstract:
This study describes a micro device integrated with multi-chamber for polymerase chain reaction (PCR) with different annealing temperatures. The device consists of the reaction polydimethylsiloxane (PDMS) chip, a cover glass chip, and is equipped with cartridge heaters, fans, and thermocouples for temperature control. In this prototype, commercial software is utilized to determine the geometric and operational parameters those are responsible for creating the denaturation, annealing, and extension temperatures within the chip. Two cartridge heaters are placed at two sides of the chip and maintained at two different temperatures to achieve a thermal gradient on the chip during the annealing step. The temperatures on the chip surface are measured via an infrared imager. Some thermocouples inserted into the reaction chambers are used to obtain the transient temperature profiles of the reaction chambers during several thermal cycles. The experimental temperatures compared to the simulated results show a similar trend. This work should be interesting to persons involved in the high-temperature based reactions and genomics or cell analysis.Keywords: Polymerase chain reaction, thermal cycles, temperature gradient, micro-fabrication.
Digital Object Identifier (DOI): doi.org/10.5281/zenodo.1077585
Procedia APA BibTeX Chicago EndNote Harvard JSON MLA RIS XML ISO 690 PDF Downloads 1652References:
[1] R. K. Saiki, S. Scharf, F. Faloona, K.B. Mullis, G. T. Horn, H. A. Erlich, and N. Arnheim, "Enzymatic amplification of beta-globin genomic sequences and restriction site analysis for diagnosis of sickle cell anemia," Science, vol. 230, pp. 1950-1954, 1985.
[2] M. A. Northrup, M. T. Ching, R. M. White, and R. T. Wltson, "DNA amplification in a microfabricated reaction chamber," in Proceeding of the 7th international conference of solid state sensors and actuators, pp 924-926, 1993.
[3] Q. Xiang, B. Xu, R. Fu, and D. Li, "Real time PCR on disposable PDMS chip with a miniaturized thermal cycler," Biomed. Microdevices, vol. 7, pp. 273-279, 2005.
[4] T. Ohashi, H. Kuyama, N. Hanafusa, and Y. Togawa, "A simple device using magnetic transportation for droplet-based PCR," Biomed. Microdevices, vol. 9, pp. 695-702, 2007.
[5] Y. Sun, N. T. Nguyen, and Y. C. Kwok, "High-throughput polymerase chain reaction in parallel circular loops using magnetic actuation," Anal. Chem., vol. 80, pp. 6127-6130, 2008.
[6] N. Ramalingam, Z. Rui, H. B. Liu, C. C. Dai, R. Kaushik, B. Ratnaharika, and H. Q. Gong, "Real-time PCR-based microfluidic array chip for simultaneous detection of multiple waterborne pathogens," Sensor. Actuat. B-Chem., vol. 145, pp. 543-552, 2010.
[7] C. Zhang, and D. Xing, "Microfluidic gradient PCR (MG-PCR): a new method for microfluidic DNA amplification," Biomed. Microdevices, vol. 12, pp. 1-12, 2010.
[8] E. Wulff-Burchfield, W. A. Schell, A. E. Eckhardt, M. G. Pollack, Z. Hua, J. L. Rouse, V. K. Pamula, V. Srinivasan, J. L. Benton, B. D. Alexander, D. A. Wilfret, M. Kraft, C. B. Cairns, J. R. Perfect, and T. G. Mitchell, "Microfluidic platform versus conventional real-time polymerase chain reaction for the detection of Mycoplasma pneumoniae in respiratory specimens," Diag. Micr. Infec. Dis., vol. 67, pp. 22-29, 2010.
[9] D. Sugumar, L. X. Kong, A. Ismail, M. Ravichandran, and L. S. Yin, "Rapid multi sample DNA amplification using rotary-linear polymerase chain reaction device (PCRDisc)," Biomicrofluidics, vol. 6, 014119, 2012.
[10] D. Liu, G. Liang, X. Lei, B. Chen, W. Wang, and X. Zhou, "Highly efficient capillary polymerase chain reaction using an oscillation droplet microreactor," Anal. Chim. Acta, vol. 718, pp. 58-63, 2012.
[11] J. P. Holman, Heat Transfer, 10th ed. McGraw-Hill, New York, 2009.
[12] T. J. Barth, and D. C. Jespersen, "The design and application of upwind schemes on unstructured meshes," in 27th Aerospace Sciences Meeting and Exhibit, AIAA-89-0366, 1989.
[13] L. H. Howell, R. B. Pember, P. Colella, J. P. Jessee, and W. A. Fiveland, "A conservative adaptive-mesh algorithm for unsteady, combined-mode heat transfer using the discrete ordinates method," Numer. Heat Tr. B-Fund., vol. 35, pp. 407-430, 1999.
[14] J. J. Chen, C. H. Chen, and S. R. Shie, "Optimal designs of staggered Dean vortex micromixers," Int. J. Mol. Sci., vol. 12, pp. 3500-3524, 2011.