Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 87758
Relation between Initial Stability of the Dental Implant and Bone-Implant Contact Level
Authors: Jui-Ting Hsu, Heng-Li Huang, Ming-Tzu Tsai, Kuo-Chih Su, Lih-Jyh Fuh
Abstract:
The objectives of this study were to measure the initial stability of the dental implant (ISQ and PTV) in the artificial foam bone block with three different quality levels. In addition, the 3D bone to implant contact percentage (BIC%) was measured based on the micro-computed tomography images. Furthermore, the relation between the initial stability of dental implant (ISQ and PTV) and BIC% were calculated. The experimental results indicated that enhanced the material property of the artificial foam bone increased the initial stability of the dental implant. The Pearson’s correlation coefficient between the BIC% and the two approaches (ISQ and PTV) were 0.652 and 0.745.Keywords: dental implant, implant stability quotient, peak insertion torque, bone-implant contact, micro-computed tomography
Procedia PDF Downloads 581