Coded Transmission in Synthetic Transmit Aperture Ultrasound Imaging Method
Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 33122
Coded Transmission in Synthetic Transmit Aperture Ultrasound Imaging Method

Authors: Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki, Marcin Lewandowski

Abstract:

The paper presents the study of synthetic transmit aperture method applying the Golay coded transmission for medical ultrasound imaging. Longer coded excitation allows to increase the total energy of the transmitted signal without increasing the peak pressure. Signal-to-noise ratio and penetration depth are improved maintaining high ultrasound image resolution. In the work the 128-element linear transducer array with 0.3 mm inter-element spacing excited by one cycle and the 8 and 16-bit Golay coded sequences at nominal frequencies 4 MHz was used. Single element transmission aperture was used to generate a spherical wave covering the full image region and all the elements received the echo signals. The comparison of 2D ultrasound images of the wire phantom as well as of the tissue mimicking phantom is presented to demonstrate the benefits of the coded transmission. The results were obtained using the synthetic aperture algorithm with transmit and receive signals correction based on a single element directivity function.

Keywords: Golay coded sequences, radiation pattern, synthetic aperture, ultrasound imaging.

Digital Object Identifier (DOI): doi.org/10.5281/zenodo.1074385

Procedia APA BibTeX Chicago EndNote Harvard JSON MLA RIS XML ISO 690 PDF Downloads 2135

References:


[1] Y. Tasinkevych, A. Nowicki, I. Trots, "Element directivity influence in the synthetic focusing algorithm for ultrasound imaging," in Proc. 57th Open Seminar on Acoustics, Gliwice, Poland, 2010, pp. 197-200.
[2] J.A. Jensen, "Linear description of ultrasound imaging systems," Note for the International Summer School on Advanced Ultrasound Imaging, Technical University of Denmark, June 10, 1999.
[3] G.E. Trahey, L.F. Nock, "Synthetic receive aperture imaging with phase correction for motion and for tissue ihomogeneities ÔÇö part I: Basic principles," IEEE Trans. Ultrason. Ferroelec. Freq. Contr., vol. 39, no. 4, pp. 489-495, 1992.
[4] I. Trots, A. Nowicki, M. Lewandowski, "Synthetic transmit aperture in ultrasound imaging," Archives of Acoustics, vol. 43, no. 4, pp. 685-695, 2009.
[5] Y. Tasinkevych, I. Trots, A. Nowicki, P.A. Lewin, "Modified synthetic transmit aperture algorithm for ultrasound imaging," Ultrasonics, vol. 52, no. 2, pp. 333-342, 2012.
[6] A.R. Selfridge, G.S. Kino, B.T. Khuri-Yakub, "A theory for the radiation pattern of a narrow-strip acoustic transducer," Appl. Phys. Lett., vol. 37, no. 1, pp. 35-36, 1980.
[7] M.J.E. Golay, "Complementary series," IRE Tran. Inf. Theory, vol. 7, pp. 82-87, 1961.
[8] E.J. Danicki, "Complementary code realization based on surface acoustic waves," Bulletin of Military Technical Academy, vol. XXIII, no. 1, pp. 53-56, 1974.
[9] I. Trots, A. Nowicki, W. Secomski, J. Litniewski, "Golay sequences - side-lobe canceling codes for ultrasonography," Archives of Acoustics, vol. 29, no. 1, pp. 87-97, 2004.
[10] A. Nowicki, Z. Klimonda, M. Lewandowski, J. Litniewski, P.A. Lewin, I. Trots, "Direct and post-compressed sound fields for different coded excitation," Acoustical Imaging, vol. 28, pp. 399-407, 2007.
[11] M. Xu, L.V. Wang, "Analytic explanation of spatial resolution related to bandwidth and detector aperture size in thermoacoustic or photoacoustic reconstruction," Phys. Rev. E, vol. 67, no. 5, pp. 1-15, 2003.