Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 33122
Directionally-Sensitive Personal Wearable Radiation Dosimeter
Authors: Hai Huu Le, Paul Junor, Moshi Geso, Graeme O’Keefe
Abstract:
In this paper, the authors propose a personal wearable directionally-sensitive radiation dosimeter using multiple semiconductor CdZnTe detectors. The proposed dosimeter not only measures the real-time dose rate but also provide the direction of the radioactive source. A linear relationship between radioactive source direction and the radiation intensity measured by each detectors is established and an equation to determine the source direction is derived by the authors. The efficiency and accuracy of the proposed dosimeter is verified by simulation using Geant4 package. Results have indicated that in a measurement duration of about 7 seconds, the proposed dosimeter was able to estimate the direction of a 10μCi 137/55Cs radioactive source to within 2 degrees.Keywords: Dose rate, Geant4 package, radiation detectors, radioactive source direction.
Digital Object Identifier (DOI): doi.org/10.5281/zenodo.1130461
Procedia APA BibTeX Chicago EndNote Harvard JSON MLA RIS XML ISO 690 PDF Downloads 1199References:
[1] I. Orion, A. Pernick, D. Ilzycer, and H. Zafrir, Locating gamma radiation source by self collimating BGO detector system, Israel Nuclear Society, Conference of the Israel Nuclear Societies, 9-10 Dec 1996, 1-4.
[2] J.L. Matteson, M.A. Capote, R.T. Skelton, B. Greg, S. Edwin, R. Richard, H. George, G. Tom and P. Michael, A Directional Gamma Radiation Spectrometer Based on Pixelated CZT Arrays and Coded Mask Apertures, Nuclear Science Symposium Conference Record, 2006, 87-92.
[3] J. Uher, C. Frojdh, S. Pospisil, G. Thungstrom and Z. Vykydal, Directional radiation detector, Nuclear Science Symposium Conference Record, 2007, 1162-1165.
[4] M.Willis, S. Skutnik and H. Hall, Detection and positioning of radioactive sources using a four-detector response algorithm, Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014, 445-452.
[5] Geant4 Collaboration, Geant4 User’s Guide for Application Developers, http://geant4.web.cern.ch/geant4/UserDocumentation/UsersGuides/ForApplicationDeveloper/ BackupVersions/V10.2/fo/BookForAppliDev.pdf, accessed on 20/05/2016.
[6] S.D. Sordo, L. Abbene, E. Caroli, A.M. Mancini, A. Zappettini, and P. Ubertini, Progress in the Development of CdTe and CdZnTe Semiconductor Radiation Detectors for Astrophysical and Medical Applications, Sensors 9, doi:10.3390/s90503491, 2009, 3491-3526