An in vitro study was carried out to evaluate the feasibility of small field of view (FOV) cone beam computed tomography (CBCT) in determining endodontic working length. The objectives were to determine the accuracy of CBCT in measuring the estimated preoperative working lengths (EPWL), endodontic working lengths (EWL) and file lengths. Access cavities were prepared in 27 molars. For each root canal, the baseline electronic working length was determined using an EAL (Raypex 5). The teeth were then divided into overextended, non-modified and underextended groups and the lengths were adjusted accordingly. Imaging and measurements were made using the respective software of the RVG (Kodak RVG 6100) and CBCT units (Kodak 9000 3D). Root apices were then shaved and the apical constrictions viewed under magnification to measure the control working lengths. The paired t-test showed a statistically significant difference between CBCT EPWL and control length but the difference was too small to be clinically significant. From the Bland Altman analysis, the CBCT method had the widest range of 95% limits of agreement, reflecting its greater potential of error. In measuring file lengths, RVG had a bigger window of 95% limits of agreement compared to CBCT. Conclusions: (1) The clinically insignificant underestimation of the preoperative working length using small FOV CBCT showed that it is acceptable for use in the estimation of preoperative working length. (2) Small FOV CBCT may be used in working length determination but it is not as accurate as the currently practiced method of using the EAL. (3) It is also more accurate than RVG in measuring file lengths.

Keywords: Accuracy, CBCT, endodontic, measurement.

Digital Object Identifier (DOI): doi.org/10.5281/zenodo.1123887

Procedia APA BibTeX Chicago EndNote Harvard JSON MLA RIS XML ISO 690 PDF Downloads 1624

References:


[1] Vertucci, “Root canal anatomy of the human permanent teeth,” Oral Surg Oral Med Oral Pathol, vol. 58, pp. 589-599, 1984.
[2] C. L. Burger, T. O. Mork, J. W. Hutter, B. Nicoll, “Direct digital radiography versus conventional radiography for estimation of canal length in curved canals,” J Endod, vol. 25, no. 4, pp. 260-264, 1999.
[3] C.B. Williams, A.P. Joyce, S. Roberts, “A Comparison between in vivo radiographic working length determination and measurement after extraction,” J Endod, vol. 32, no.7, pp 624-627, 2006.
[4] S.C. White and M.J. Pharoah, In Oral Radiology: principles and interpretation, 6th edition. Mosby Elsevier, St. Louis, Missouri, 2009, pp. 225-243.
[5] A. ElAyouti, R. Weiger, C. Löst, “The ability of root ZX apex locator to reduce the frequency of overestimated radiographic working length,” J Endod, vol. 28, no. 2, pp. 116-119, 2002.
[6] R. J. Czerw, M.S. Fulkerson, J. C. Donnelly, J. O. Walmann, “In vitro evaluation of the accuracy of several electronic apex locators,” J Endod, vol. 21, no. 11, pp. 572-575, 1995.
[7] Lucena-Martín C., Robles-Gijón V., Ferrer-Luque C.M., Mondelo J.M. (2004). In vitro evaluation of the accuracy of three apex locators. J Endod, 30 (4), 231-233
[8] F. Goldberg, B. B. Marroquín, S. Frajlich, “In vitro evaluation of the ability of three apex locators to determine the working length during retreatment,” J Endod, vol. 31, no. 9, pp. 676-678, 2005.
[9] F. L. C. D́Assunção, D. S. D. Albuquerque, L. C. D. Q. Ferreira, L. C. Ferreira, “The ability of two apex locators to locate the apical foramen: an in vitro study,” J Endod, vol. 32, no.6, pp. 560-562, 2006.
[10] F. L. C. D’Assunção, D. S. D. Albuquerque, J. R. Salazar-Silva, D. Q. Ferreira, P. M. Bezerrz, “The accuracy of root canal measurements using the mini apex locator and Root ZX-II: an evaluation in vitro,” Oral Surg Oral Med Oral Pathol Oral Radiol Endod, vol. 104, no.3, e50-e53, 2007.
[11] R. A. Bernardes, M. A. H. Duarte, B. C. Vasconcelos, I. G. Moraes, N. Bernardinelli, R. B. Garcia, J. V. Baldi, F. R. Victorino, C. M. Bramante, “Evaluation of precision of length determination with 3 electronic apex locator: Root ZX, Elements Diagnostic Unit and Apex locator and RomiAPEX D-30,” Oral Surg Oral Med Oral Pathol Oral Radiol Endod, vol. 104, no.4, e91-e94, 2007
[12] T. P. Cotton, T. M. Geisler, D. T. Holden, S. A. Schwartz, W. G. Schindler, “Endodontic applications of Cone-Beam Volumetric Tomography,” J Endod, vol. 33, no.9, pp. 1121-1132, 2007.
[13] M. K. Nair and U. P. Nair, “Digital and advanced imaging in endodontics: a review,” J Endod, vol. 33, no. 1-6, 2007a.
[14] C. Estrela, M. R. Bueno, C. R. Leles, B. Azevedo, J. R. Azevedo, “Accuracy of cone-beam computed tomography and panoramic and periapical radiography for detection of apical periodontitis,” J Endod, vol. 34, pp. 273-279, 2008.
[15] Y. Kuttler, “Microscopic investigations of root apexes,” J Am Dent Assoc, vol. 50, p.544, 1955.
[16] M. K. Wu, P. R. Wesselink, R. E. Walton, “Apical terminus location of root canal treatment procedures,” Oral Surg Oral Med Oral Pathol Oral Radiol Endod, vol. 89, pp. 99-103, 2000.
[17] M. A. Kim-Park, L. W. Baughan, Hartwell, “Working length determination in palatal roots of maxillary molars,” J Endod, vol. 29, no.1, pp. 58-61, 2003.
[18] J. B. Ludlow, W. S. Laster, M. See, L. J. Bailey, H. G. Hershey, “Accuracy of measurements of mandibular anatomy in cone beam computed tomography images,” Oral Surg Oral Med Oral Pathol Oral Radiol Endod, vol. 103, pp. 534–42, 2007.
[19] R. Marmulla, R. Wortche, J. Muhling, S. Hassfeld, “Geometric accuracy of the NewTom 9000 Cone Beam CT,” Dentomamax Radiol, vol. 34, pp. 28-31, 2005.
[20] K. Tsutsumi, T. Chikui, K. Okamura, K. Yoshiura, “Accuracy of linear measurement and the measurement limits of thin objects with cone beam computed tomography: effects of measurement directions and of phatom locations in the fields of view,” Int J Oral Maxillofac Implants, vol. 26, pp. 91-100, 2011.
[21] M. O. Lagravere, J. Carey, R. W. Toogood, P. W. Major, “Three dimensional accuracy of measurements made with software on cone-beam computed tomography images,” Am J Orth Dentofac Orthop, vol. 134, no.1, pp.112-116, 2008.