The Effect of Screw Parameters on Pullout Strength of Screw Fixation in Cervical Spine
Authors: S. Ritddech, P. Aroonjarattham, K. Aroonjarattham
Abstract:
The pullout strength had an effect on the stability of plate screw fixation when inserted in the cervical spine. Nine different titanium alloy bone screws were used to test the pullout strength through finite element analysis. The result showed that the Moss Miami I can bear the highest pullout force at 1,075 N, which causes the maximum von Mises stress at 858.87 MPa, a value over the yield strength of titanium. The bone screw should have large outer diameter, core diameter and proximal root radius to increase the pullout strength.
Keywords: Pullout strength, Screw parameter, Cervical spine, Finite element analysis.
Digital Object Identifier (DOI): doi.org/10.5281/zenodo.1099668
Procedia APA BibTeX Chicago EndNote Harvard JSON MLA RIS XML ISO 690 PDF Downloads 3158References:
[1] T.A. Mattei, M.S. Meneses, J.B. Milano and R. Ramina. “Free-hand” technique for thoracolumbar pedicle screw instrumentation: critical appraisal of current “state of art” Neurol India. 2009 vol. 57(6), pp. 715- 721.
[2] J.D. Thompson, J.B. Benjamin and J.A. Szivek “Pullout strengths of cannulated and non-cannulated cancellous bone screws” Clinical Orthopaedics and Related Research 1997 vol. 341, pp. 241-249.
[3] T. Hirano, K. Hasegawa, H.E. Takahashi, S. Uchiyama, T. Hara, T. Washio, T. Sugiura, M. Yokaichiya and M. Ikeda. “Structural characteristics of the pedicle and its role in screw stability” Spine 1997 vol. 22, pp. 2504-2509
[4] T.H. Lim, H.S. An, C. Evanich, K.Y. Hasanoglu, L. Mcgrady and C.R. Wilson. “Strength of anterior vertebral screw fixation in relationship to bone mineral density” Journal of Spinal Disorder 1995 vol. 8, pp. 121- 125.
[5] Q.H. Zhang, S.H. Tan and S.M. Chou. “Investigation of fixation screw pull-out strength on human spine” J. Biomechanics 2004 vol. 37, pp. 479-485.
[6] C.H. Ching, K.C. Ching, L.W. Jaw, M.H. Sheng, T.T. Ying and L. Jinn. “Increase of pullout strength of spinal pedicle screws with conical core: biomechanical tests and finite element analyse” J. Orthopaedic Research 2005 vol. 23, pp. 788-794.
[7] E.E. Karaikovic, S. Kunakornsawat, M.D. Daubs, R. Madsen and R.W. Jr. Gaines. “Surgical anatomy of the cervical pedicles (landmarks for posterior cervical pedicle entrance localization)” J. Spinal Disord 2000 vol. 13, pp. 63-72.
[8] A. Perez, A. Mahar, C. Negus, P. Newton and T. Impelluso. “A computational evaluation of the effect of intramedullary nail material properties on the stabilization of simulated femoral shaft fractures” J. Med Eng & Phy 2007 vol. 20, pp. 755-760.
[9] A. Ramos and J.A. Simoes. “Tetrahedral versus hexahedral finite elements in numerical modeling of the proximal femur” J. Med Eng&Phy 2006 vol. 28, pp. 916-924.
[10] Z. Paul, B. David, E.G. Allan and P. Marcello. “Nothching of the Anterior Femoral Cortex during Total Knee Arthoplasty” J. Arthroplasty 2006 vol. 21, pp. 737-743.
[11] S.L. Chen, R.M. Lin and C.H. Chang. “Biomechanical investigation of pedicle screw-vertebrae complex: a finite element approach using bonded and contact interface conditions” Med Eng & Phy 2003 vol. 25, pp. 275-282.
[12] J. Black and G. Hastings. 1998 “Handbook of biomaterials properties” Chapman & Hall, UK.
[13] Q.H. Zhang, S.H. Tan and S.M. Chou. “Effects of bone materials on the screw pull-out strength in human spine” Med Eng & Phy 2006 vol. 28, pp. 795-801.
[14] J.R. Chapman, R.M. Harrington, K.M. Lee, P.A. Anderson, A.F. Tencer and D. Kowalski. “Factors affecting the pullout strength of cancellous bone screws” J. Biomech Eng. 1996 vol. 118, pp. 391-398.
[15] A. Ono, M.D. Brown, L.L. Latta, E.L. Milne and D.C. Holmes. ''Triangulated pedicle screw construct technique and pull-out strength of conical and cylindrical screw'' J. Spinal Disord 2001 vol. 14, pp. 323- 329.
[16] B.B. Abshire, R.F. McLain, A. Valdevit and H.E. Kambic. ''Characteristics of pullout failure in conical and cylindrical pedicle screws after full insertion and back-out'' J. Spine 2001 vol. 1, pp. 408- 414.
[17] T. Gausepohl, R. Mohring, D. Penning and J. Koebke. ''Fine thread versus coarse thread: a comparison of the maximum holding power'' Injury2001 vol. 32, pp. 1-7.
[18] C. Altieri, J. Flores, V. Gonzalez and A. Rodriguez. “Biomechanics of orthopaedic fixations” 2003 Mechanics of Materials I. University of Puerto Rico.