Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 2
Search results for: P. Aroonjarattham
2 The Effect of Screw Parameters on Pullout Strength of Screw Fixation in Cervical Spine
Authors: S. Ritddech, P. Aroonjarattham, K. Aroonjarattham
Abstract:
The pullout strength had an effect on the stability of plate screw fixation when inserted in the cervical spine. Nine different titanium alloy bone screws were used to test the pullout strength through finite element analysis. The result showed that the Moss Miami I can bear the highest pullout force at 1,075 N, which causes the maximum von Mises stress at 858.87 MPa, a value over the yield strength of titanium. The bone screw should have large outer diameter, core diameter and proximal root radius to increase the pullout strength.Keywords: pullout strength, screw parameter, cervical spine, finite element analysis
Procedia PDF Downloads 2921 Static and Dynamic Load on Hip Contact of Hip Prosthesis and Thai Femoral Bones
Authors: K. Chalernpon, P. Aroonjarattham, K. Aroonjarattham
Abstract:
Total hip replacement had been one of the most successful operations in hip arthritis surgery. The purpose of this research had been to develop a dynamic hip contact of Thai femoral bone to analyze the stress distribution on the implant and the strain distribution on the bone model under daily activities and compared with the static load simulation. The results showed the different of maximum von Mises stress 0.14 percent under walking and 0.03 percent under climbing stair condition and the different of equivalent total strain 0.52 percent under walking and 0.05 percent under climbing stair condition. The muscular forces should be evaluated with dynamic condition to reduce the maximum von Mises stress and equivalent total strain.Keywords: dynamic loading, static load, hip prosthesis, Thai femur, femoral bone, finite element analysis
Procedia PDF Downloads 349