Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 2

flow meter Related Abstracts

2 Flow Measurement Using Magnetic Meters in Large Underground Cooling Water Pipelines

Authors: Humanyun Zahir, Irtsam Ghazi

Abstract:

This report outlines the basic installation and operation of magnetic inductive flow velocity sensors on large underground cooling water pipelines. Research on the effects of cathodic protection as well as into other factors that might influence the overall performance of the meter are presented in this paper. The experiments were carried out on an immersion type magnetic meter specially used for flow measurement of cooling water pipeline. An attempt has been made in this paper to outline guidelines that can ensure accurate measurement related to immersion type magnetic meters on underground pipelines.

Keywords: Immersion, electrodes, anode, cathode, grounding, flange, magnetic induction, cathodic protection, flow meter, Faraday's law, plant information management system

Procedia PDF Downloads 283
1 A Measurement Device of Condensing Flow Rate, an Order of MilliGrams per Second

Authors: Hee Joon Lee

Abstract:

There are many difficulties in measuring a small flow rate of an order of milli grams per minute (LPM) or less using a conventional flowmeter. Therefore, a flow meter with minimal loss and based on a new concept was designed as part of this paper. A chamber was manufactured with a level transmitter and an on-off control valve. When the level of the collected condensed water reaches the top of the chamber, the valve opens to allow the collected water to drain back into the tank. To allow the water to continue to drain when the signal is lost, the valve is held open for a few seconds by a time delay switch and then closed. After an examination, the condensing flow rate was successfully measured with the uncertainty of ±5.7% of the full scale for the chamber.

Keywords: Condensation, flow meter, chamber, milli-grams

Procedia PDF Downloads 147