	%X This study aimed to present the mechanical
performance evaluation of the dynamic hip screw (DHS) for
trochanteric fracture by means of finite element method. The
analyses were performed based on stainless steel and titanium
implant material definitions at various stages of bone healing and
including implant removal. The assessment of the mechanical
performance used two parameters, von Mises stress to evaluate the
strength of bone and implant and elastic strain to evaluate fracture
stability. The results show several critical aspects of dynamic hip
screw for trochanteric fracture stabilization. In the initial stage of
bone healing process, partial weight bearing should be applied to
avoid the implant failure. In the late stage of bone healing, stainless
steel implant should be removed.