WASET
	%0 Journal Article
	%A Hakan Işık and  Esra Saraçoğlu
	%D 2010
	%J International Journal of Mechanical and Mechatronics Engineering
	%B World Academy of Science, Engineering and Technology
	%I Open Science Index 44, 2010
	%T Comparison of Proportional Control and Fuzzy Logic Control to Develop an Ideal Thermoelectric Renal Hypothermia System
	%U https://publications.waset.org/pdf/15459
	%V 44
	%X In this study, a comparison of two control methods,
Proportional Control (PC) and Fuzzy Logic Control (FLC), which
have been used to develop an ideal thermoelectric renal hypothermia
system in order to use in renal surgery, has been carried out. Since
the most important issues in long-lasting parenchymatous renal
surgery are to provide an operation medium free of blood and to
prevent renal dysfunction in the postoperative period, control of the
temperature has become very important in renal surgery. The final
product is seriously affected from the changes in temperature,
therefore, it is necessary to reach some desired temperature points
quickly and avoid large overshoot. PIC16F877 microcontroller has
been used as controller for both of these two methods. Each control
method can simply ensure extra renal hypothermia in the targeted
way. But investigation of advantages and disadvantages of every
control method to each other is aimed and carried out by the
experimental implementations. Shortly, investigation of the most
appropriate method to use for development of system and that can be
applied to people safely in the future, has been performed. In this
sense, experimental results show that fuzzy logic control gives out
more reliable responses and efficient performance.
	%P 1328 - 1335