	%T A Novel Dosimetry System for Computed Tomography using Phototransistor
	%U https://publications.waset.org/pdf/1515
	%V 56
	%X Computed tomography (CT) dosimetry normally uses
an ionization chamber 100 mm long to estimate the computed
tomography dose index (CTDI), however some reports have already
indicated that small devices could replace the long ion chamber to
improve quality assurance procedures in CT dosimetry. This paper
presents a novel dosimetry system based in a commercial
phototransistor evaluated for CT dosimetry. Three detector
configurations were developed for this system: with a single, two and
four devices. Dose profile measurements were obtained with them
and their angular response were evaluated. The results showed that
the novel dosimetry system with the phototransistor could be an
alternative for CT dosimetry. It allows to obtain the CT dose profile
in details and also to estimate the CTDI in longer length than the
100 mm pencil chamber. The angular response showed that the one
device detector configuration is the most adequate among the three
configurations analyzed in this study.
	%P 327 - 330