%0 Journal Article %A C. M. M. Paschoal and D. do N. Souza and L. A. P. Santos %D 2011 %J International Journal of Biomedical and Biological Engineering %B World Academy of Science, Engineering and Technology %I Open Science Index 56, 2011 %T Characterization of Three Photodetector Types for Computed Tomography Dosimetry %U https://publications.waset.org/pdf/13680 %V 56 %X In this study three commercial semiconductor devices were characterized in the laboratory for computed tomography dosimetry: one photodiode and two phototransistors. It was evaluated four responses to the irradiation: dose linearity, energy dependence, angular dependence and loss of sensitivity after X ray exposure. The results showed that the three devices have proportional response with the air kerma; the energy dependence displayed for each device suggests that some calibration factors would be applied for each one; the angular dependence showed a similar pattern among the three electronic components. In respect to the fourth parameter analyzed, one phototransistor has the highest sensitivity however it also showed the greatest loss of sensitivity with the accumulated dose. The photodiode was the device with the smaller sensitivity to radiation, on the other hand, the loss of sensitivity after irradiation is negligible. Since high accuracy is a desired feature for a dosimeter, the photodiode can be the most suitable of the three devices for dosimetry in tomography. The phototransistors can also be used for CT dosimetry, however it would be necessary a correction factor due to loss of sensitivity with accumulated dose. %P 297 - 300