Calibration of Syringe Pumps Using Interferometry and Optical Methods
Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 32804
Calibration of Syringe Pumps Using Interferometry and Optical Methods

Authors: E. Batista, R. Mendes, A. Furtado, M. C. Ferreira, I. Godinho, J. A. Sousa, M. Alvares, R. Martins

Abstract:

Syringe pumps are commonly used for drug delivery in hospitals and clinical environments. These instruments are critical in neonatology and oncology, where any variation in the flow rate and drug dosing quantity can lead to severe incidents and even death of the patient. Therefore it is very important to determine the accuracy and precision of these devices using the suitable calibration methods. The Volume Laboratory of the Portuguese Institute for Quality (LVC/IPQ) uses two different methods to calibrate syringe pumps from 16 nL/min up to 20 mL/min. The Interferometric method uses an interferometer to monitor the distance travelled by a pusher block of the syringe pump in order to determine the flow rate. Therefore, knowing the internal diameter of the syringe with very high precision, the travelled distance, and the time needed for that travelled distance, it was possible to calculate the flow rate of the fluid inside the syringe and its uncertainty. As an alternative to the gravimetric and the interferometric method, a methodology based on the application of optical technology was also developed to measure flow rates. Mainly this method relies on measuring the increase of volume of a drop over time. The objective of this work is to compare the results of the calibration of two syringe pumps using the different methodologies described above. The obtained results were consistent for the three methods used. The uncertainties values were very similar for all the three methods, being higher for the optical drop method due to setup limitations.

Keywords: Calibration, interferometry, syringe pump, optical method, uncertainty.

Procedia APA BibTeX Chicago EndNote Harvard JSON MLA RIS XML ISO 690 PDF Downloads 707

References:


[1] Bissig, H., Petter, H.T., Lucas, P., Batista, E. et al., Primary standards for measuring flow rates from 100 nl/min to 1 ml/min – gravimetric principle, Biomedical Engineering, 2015, 60 (4); 301-316
[2] Batista, E. et al., Comparison of infusion pumps calibration methods, Meas. Sci. Technol. 2017, 28
[3] Batista, E. et al., Assessment of drug delivery devices, Biomed. Eng. 2015, 60 (4); 347–57
[4] Batista E, Almeida N, Godinho I and Filipe E, Uncertainty Calculation in Gravimetric Microflow Measurements, AMCTM X, 2015, vol 86, pp 98–104 (London: World Scientific)
[5] Batista E., Alves e Sousa J., Cardoso S., Silvério V. Experimental testing for metrological traceability and accuracy of liquid microflows and microfluidics, Flow Measurement and Instrumentation, Volume 71, March 2020, 101691
[6] MeDD II- Metrology for drug delivery, Publishable summary at www.drugmetrology.com
[7] Batista E., et al., Development of an experimental setup for microflow measurement using interferometry, Flow Measurement and Instrumentation, Volume 75 (2020) 101789.
[8] BIPM, IEC, IFCC, ILAC, ISO, IUPAP and OIML, “Evaluation of measurement data – Supplement 1 to the Guide to the Expression of Uncertainty in Measurement – Propagation of distributions using a Monte Carlo method, Joint Committee for Guides in Metrology, JCGM 101, 2008.
[9] Batista, E, Sousa, J A, Martins, R, Calibration of insulin pumps, Journal of Diabetes and Treatment, Volume 4; Issue 03, 2019
[10] www.bipm.org, accessed on 2020-10-02