WASET
	@article{(Open Science Index):https://publications.waset.org/pdf/10001587,
	  title     = {The Effect of Choke on the Efficiency of Coaxial Antenna for Percutaneous Microwave Coagulation Therapy for Hepatic Tumor},
	  author    = {Surita Maini},
	  country	= {},
	  institution	= {},
	  abstract     = {There are many perceived advantages of microwave
ablation have driven researchers to develop innovative antennas to
effectively treat deep-seated, non-resectable hepatic tumors. In this
paper a coaxial antenna with a miniaturized sleeve choke has been
discussed for microwave interstitial ablation therapy, in order to
reduce backward heating effects irrespective of the insertion depth
into the tissue. Two dimensional Finite Element Method (FEM) is
used to simulate and measure the results of miniaturized sleeve choke
antenna. This paper emphasizes the importance of factors that can
affect simulation accuracy, which include mesh resolution, surface
heating and reflection coefficient. Quarter wavelength choke
effectiveness has been discussed by comparing it with the unchoked
antenna with same dimensions.},
	    journal   = {International Journal of Biomedical and Biological Engineering},
	  volume    = {9},
	  number    = {7},
	  year      = {2015},
	  pages     = {541 - 545},
	  ee        = {https://publications.waset.org/pdf/10001587},
	  url   	= {https://publications.waset.org/vol/103},
	  bibsource = {https://publications.waset.org/},
	  issn  	= {eISSN: 1307-6892},
	  publisher = {World Academy of Science, Engineering and Technology},
	  index 	= {Open Science Index 103, 2015},
	}