Radiochemical Purity of 68Ga-BCA-Peptides: Separation of All 68Ga Species with a Single iTLC Strip
Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 84079
Radiochemical Purity of 68Ga-BCA-Peptides: Separation of All 68Ga Species with a Single iTLC Strip

Authors: Anton A. Larenkov, Alesya Ya Maruk

Abstract:

In the present study, highly effective iTLC single strip method for the determination of radiochemical purity (RCP) of 68Ga-BCA-peptides was developed (with no double-developing, changing of eluents or other additional manipulation). In this method iTLC-SG strips and commonly used eluent TFAaq. (3-5 % (v/v)) are used. The method allows determining each of the key radiochemical forms of 68Ga (colloidal, bound, ionic) separately with the peaks separation being no less than 4 σ. Rf = 0.0-0.1 for 68Ga-colloid; Rf = 0.5-0.6 for 68Ga-BCA-peptides; Rf = 0.9-1.0 for ionic 68Ga. The method is simple and fast: For developing length of 75 mm only 4-6 min is required (versus 18-20 min for pharmacopoeial method). The method has been tested on various compounds (including 68Ga-DOTA-TOC, 68Ga-DOTA-TATE, 68Ga-NODAGA-RGD2 etc.). The cross-validation work for every specific form of 68Ga showed good correlation between method developed and control (pharmacopoeial) methods. The method can become convenient and much more informative replacement for pharmacopoeial methods, including HPLC.

Keywords: DOTA-TATE, 68Ga, quality control, radiochemical purity, radiopharmaceuticals, TLC

Procedia PDF Downloads 257