	%T A Novel Method Based on Monte Carlo for Simulation of Variable Resolution X-ray CT Scanner: Measurement of System Presampling MTF
	%U https://publications.waset.org/pdf/8450
	%V 44
	%X The purpose of this work is measurement of the
system presampling MTF of a variable resolution x-ray (VRX) CT
scanner. In this paper, we used the parameters of an actual VRX CT
scanner for simulation and study of effect of different focal spot sizes
on system presampling MTF by Monte Carlo method (GATE
simulation software). Focal spot size of 0.6 mm limited the spatial
resolution of the system to 5.5 cy/mm at incident angles of below 17º
for cell#1. By focal spot size of 0.3 mm the spatial resolution
increased up to 11 cy/mm and the limiting effect of focal spot size
appeared at incident angles of below 9º. The focal spot size of 0.3
mm could improve the spatial resolution to some extent but because
of magnification non-uniformity, there is a 10 cy/mm difference
between spatial resolution of cell#1 and cell#256. The focal spot size
of 0.1 mm acted as an ideal point source for this system. The spatial
resolution increased to more than 35 cy/mm and at all incident angles
the spatial resolution was a function of incident angle. By the way
focal spot size of 0.1 mm minimized the effect of magnification nonuniformity.
	%P 355 - 359