	%T Kerma Profile Measurements in CT Chest Scans– a Comparison of Methodologies
	%U https://publications.waset.org/pdf/7968
	%V 56
	%X The Brazilian legislation has only established
diagnostic reference levels (DRLs) in terms of Multiple Scan
Average Dose (MSAD) as a quality control parameter for computed
tomography (CT) scanners. Compliance with DRLs can be verified
by measuring the Computed Tomography Kerma Index (Ca,100) with
a pencil ionization chamber or by obtaining the kerma distribution in
CT scans with radiochromic films or rod shape lithium fluoride
termoluminescent dosimeters (TLD-100). TL dosimeters were used
to record kerma profiles and to determine MSAD values of a Bright
Speed model GE CT scanner. Measurements were done with
radiochromic films and TL dosimeters distributed in cylinders
positioned in the center and in four peripheral bores of a standard
polymethylmethacrylate (PMMA) body CT dosimetry phantom.
Irradiations were done using a protocol for adult chest. The
maximum values were found at the midpoint of the longitudinal axis.
The MSAD values obtained with three dosimetric techniques were
compared.