	%X The passive electrical properties of a tissue depends
on the intrinsic constituents and its structure, therefore by measuring
the complex electrical impedance of the tissue it might be possible to
obtain indicators of the tissue state or physiological activity [1].
Complete bio-impedance information relative to physiology and
pathology of a human body and functional states of the body tissue or
organs can be extracted by using a technique containing a fourelectrode
measurement setup. This work presents the estimation
measurement setup based on the four-electrode technique. First, the
complex impedance is estimated by three different estimation
techniques: Fourier, Sine Correlation and Digital De-convolution and
then estimation errors for the magnitude, phase, reactance and
resistance are calculated and analyzed for different levels of
disturbances in the observations. The absolute values of relative
errors are plotted and the graphical performance of each technique is
compared.