An Ultra-Low Output Impedance Power Amplifier for Tx Array in 7-Tesla Magnetic Resonance Imaging
Commenced in January 2007
Frequency: Monthly
Edition: International
Paper Count: 32797
An Ultra-Low Output Impedance Power Amplifier for Tx Array in 7-Tesla Magnetic Resonance Imaging

Authors: Ashraf Abuelhaija, Klaus Solbach

Abstract:

In Ultra high-field MRI scanners (3T and higher), parallel RF transmission techniques using multiple RF chains with multiple transmit elements are a promising approach to overcome the high-field MRI challenges in terms of inhomogeneity in the RF magnetic field and SAR. However, mutual coupling between the transmit array elements disturbs the desirable independent control of the RF waveforms for each element. This contribution demonstrates a 18 dB improvement of decoupling (isolation) performance due to the very low output impedance of our 1 kW power amplifier.

Keywords: EM coupling, Inter-element isolation, Magnetic resonance imaging (MRI), Parallel Transmit.

Digital Object Identifier (DOI): doi.org/10.5281/zenodo.1339049

Procedia APA BibTeX Chicago EndNote Harvard JSON MLA RIS XML ISO 690 PDF Downloads 1683

References:


[1] Y. Soutome, Y. Otake, and Y. Bito, “Vertical loop decoupling method for gapped phased-array coils,” in Proc. Intl. Soc. Mag. Reson. Med, vol. 19, 2011, p. 1859.
[2] A. Abuelhaija, K. Solbach, and A. Buck, “Power amplifier for magnetic resonance imaging using unconventional cartesian feedback loop,” accepted to be published in GeMiC, 2015.
[3] N. Hollingsworth, K. Moody, J.-F. Nielsen, D. Noll, M. P. McDougall, and S. M. Wright, “Tuning ultra-low output impedance amplifiers for optimal power and decoupling in parallel transmit MRI,” in Biomedical Imaging (ISBI), 2013 IEEE 10th International Symposium on. IEEE, 2013, pp. 946–949.
[4] X. Chu, Y. Liu, J. Sabate, and Y. Zhu, “Ultra-low output impedance RF power amplifier array,” in Proc. Intl. Soc. Mag. Reson. Med, vol. 15, 2007, p. 172.
[5] K. Solbach, A. Abuelhaija, and S. Shooshtary, “Near-magnet power amplifier with built-in coil current sensing,” in 22nd Proc. Intl. Soc. MRM, p. 1287, 2014.
[6] A. Abuelhaija, K. Solbach, and S. Orzada, “Comprehensive study on coupled meandered microstrip line rf coil elements for 7-Tesla magnetic resonance imaging,” accepted to be published in EUCAP, 2015.