TY - JFULL AU - S. Botman and D. Borchevkin and V. Petrov and E. Bogdanov and M. Patrushev and N. Shusharina PY - 2015/10/ TI - Photoplethysmography-Based Device Designing for Cardiovascular System Diagnostics T2 - International Journal of Biomedical and Biological Engineering SP - 688 EP - 693 VL - 9 SN - 1307-6892 UR - https://publications.waset.org/pdf/10002239 PU - World Academy of Science, Engineering and Technology NX - Open Science Index 105, 2015 N2 - In this paper, we report the development of the device for diagnostics of cardiovascular system state and associated automated workstation for large-scale medical measurement data collection and analysis. It was shown that optimal design for the monitoring device is wristband as it represents engineering trade-off between accuracy and usability. Monitoring device is based on the infrared reflective photoplethysmographic sensor, which allows collecting multiple physiological parameters, such as heart rate and pulsing wave characteristics. Developed device uses BLE interface for medical and supplementary data transmission to the coupled mobile phone, which processes it and send it to the doctor's automated workstation. Results of this experimental model approbation confirmed the applicability of the proposed approach. ER -